The NP-30H 3-part Hematology Analyzer ,Is easy to use with color 10.4 inch LCD , one touch to start testing and friendly operation menus. With only two routine reagents and low reagent consumption for the each test , The NP-30H 3-part Hematology Analyzer guarantees accurate measurement for the low value especially PLT using advanced Sweep-Flow technology. The NP-30H 3-part Hematology Analyzer provide 21 reportable parameters, data memory up to 100,000 results, built-in thermal printer (or external printer) and support LIS connection.
Test Menu
Parameters | ||||
---|---|---|---|---|
WBC | Neu# | Lym# | Mon# | Eos# |
Bas# | Neu% | Lym% | Mon% | Eos% |
Bas% | RBC | HGB | HCT | MCV |
MCH | MCHC | RDW-SD | RDW-CV | PLT |
MPV | PDW | PCT | P-LCR | P-LCC |
4 research parameters: ALY#,ALY%,LIC#,LIC% | 3 histograms for WBC,RBC and PLT | 3 DIFF scattergrams and one BASO scattergram for WBC differention |
Specification
Principle | Flow Cytometry (FCM) + Tri-angle laser scatter+ Chemical staining method for WBC differention Impedance method for RBC and PLT test Cyanide free colorimetry for HGB test |
---|---|
Sample Mode | Whole blood, capillary whole blood and pre-diluted modes |
Printout | Internal printer, 57 mm thermal printer with various formats |
Display | 10.4 inches TFT touch screen |
Sample volume | 20µL |
Storage | Up to 50,000 records |
Throughput | Up to 60 tests per hour |
Interface | 4 USB ports,1 LAN port |
Dimension | 364mm(W)*498mm(D)*431mm(H) |
Communication | LAN port Support HL7 protocol support bi-directional LIS |
Netweight | 28kg |
Power requirement | 100V-240V~,50/60Hz |
WBC | RCB | HGB | MCV | PLT | |
---|---|---|---|---|---|
Linearity range | ≤2%(4.0-15.0x 109 /L) | ≤1.5%(3.5-6.0x 1012/L) | ≤1.5%( 110-180g/L) | ≤1 %(70-120fL) | ≤6.0%(100-149x109/L) ≤4.0%(150-500x 109 /L) |